Filter integrity test — max pressure parameter changed during execution

During post-use filter integrity testing on lot G0131A0226BBDS, the Palltronic Flowstar IV reported “Error Maximum Pressure Reached”. The programmed maximum-pressure parameter was changed during execution, outside change management. Found 35 days later through audit-trail review.

Lot

G0131A0226BBDS

Detection latency

35 days

08 MAY 2026 → 12 JUN 2026

Found by

Audit-trail review

Not caught at execution

The question that decides this case

Impact is reported as two verdicts and never collapsed into one.

What the product data says

Bubble point measured ~5550 mbar against a 3450 mbar minimum. The filter was integral — no product impact. TR-000624

What the quality system says

A controlled parameter was altered outside change management and the GMP record obscures the execution issue. DEV-02470 narrative

Why they stay separate

Collapsing the two verdicts into one is what lets a compliance failure hide behind an intact product. SOP-QM-GEN-000005

Replay warning. Evaluated against the 12JUN2026 snapshot — the date this deviation was actually triaged — this determination is blocked at the impact node: TR-000624 is not effective as of 2026-06-12, so product SISPQ impact is BLOCKED pending an SME-authored controlled assessment. Re-running against today's corpus reproduces a better-informed decision, not the original one.